SLDL2510‑2M‑P Acoustic Doppler Current Profiler (ADCP)
SLDL2510‑2M‑P Acoustic Doppler Current Profiler (ADCP) Mainly used for multi‑layer vertical flow velocity and discharge measurement.
- Product Introduction
Solidat Applied Technologies Ltd. is one of the leading manufacturers and suppliers of sldl2510‑2m‑p acoustic doppler current profiler (adcp) in China, also supports customized service. Please feel free to buy high quality products in stock here and get quotation from our factory. Good service and reasonable price are available.
SLDL2510‑2M‑P Acoustic Doppler Current Profiler (ADCP)
SLDL2510‑2M‑P Acoustic Doppler Current Profiler (ADCP) Mainly used for multi‑layer vertical flow velocity and discharge measurement.
Product Characteristics
‑ Easy‑to‑operate current profiler, capable of acquiring velocity profile, water‑level and other data parameters.
‑ Powerful digital signal‑processing technology; velocity profile supports up to 256 measurement bins.
‑ Supports output of forward / reverse flow direction.
‑ Built‑in underwater static‑pressure water‑level sensor; external water‑level sensor is also supported.
‑ Communication interfaces: RS485 (standard Modbus protocol), SDI‑12, RS‑232, RS‑422.
‑ Supports input of field parameters and irregular cross‑section diagrams for calculating water‑cross‑section area and flow discharge.
‑ Integrated attitude sensor for easy installation with minimal civil‑engineering work.
‑ Low‑power‑consumption design for long‑term field deployment.
‑ User‑friendly configuration, diagnosis and commissioning software.
‑ Explosion‑proof grade of profiler: Ex ia.
Technical Parameters
| Item | Specification |
| Model | SLDL2510‑2M‑P |
| Operating Frequency | 2000 kHz |
| Acoustic Path | Planar‑array single‑beam |
| Acoustic Wave Measurement Range | 0.06‑25 m |
| Beam Spread Angle | 1.4° |
| Maximum Ping Rate | 20 Hz |
| Blind Zone | 0.03 m |
| Max Number of Measurement Bins | 256 |
| Bin Size | 0.15‑4 m, dynamically adjustable |
| Measurement Accuracy | ±0.5%FS ±1.0 mm/s |
| Velocity Resolution | 0.001 m/s |
| Velocity Measurement Range | ±10 m/s |
| Temperature Sensor Range | ‑25℃‑75℃ |
| Temperature Accuracy | ±0.4℃ |
| Built‑in Pressure Water‑level Measurement | 0‑20 m, ±0.1%FS |
| External Water‑level Input | Supported |
| Measurement Interval | 1‑60 min adjustable |
| Data Output | Water‑level, layered velocity, average velocity, cross‑section area, discharge, etc. |
| Operating Temperature | ‑10℃‑60℃ |
| Storage Temperature | ‑20℃‑70℃ |
| Explosion‑proof Grade | Ex ib / Ex ia |
| Protection Grade | IP68 |
| Operating Voltage | DC 7‑15 V |
| Power Consumption | Sampling<1 W |
| 3‑D Gyro‑Attitude Sensor (Optional) | True north ±2°, pitch & roll ±0.5° |
| Attitude‑Sensor Accuracy | ±0.5° |
| Communication Protocol | RS485 (standard Modbus); SDI‑12; RS‑232; RS‑422 |
| Housing Material | Stainless steel / engineering plastic optional |
| Sensor Dimensions | 295×60×60 mm |
| Weight (in air) | 1.5 kg |
| Weight (in water) | 0.3 kg |
| Reliability MTBF | >25000 h |
Hot Tags: sldl2510‑2m‑p acoustic doppler current profiler (adcp), China sldl2510‑2m‑p acoustic doppler current profiler (adcp) manufacturers, suppliers, factory, Analysis System, Portable Multiparameter Water Quality Meter, Water Analysis System